Notes in 19 Breathing Circuit

To Subscribe, use this Key


Status Last Update Fields
Published 12/13/2023 What type of breathing system has no rebreathing and no reservoir?{{c1::Open}}
Published 12/13/2023 What type of breathing system has no rebreathing and a reservoir?{{c1::Semi-open}}
Published 12/13/2023 What type of breathing system has partial rebreathing and a reservoir?{{c1::Semi-closed}}
Published 12/13/2023 What type of breathing system has complete rebreathing and a reservoir?{{c1::Closed}}
Published 12/13/2023 Open delivery of oxygen (e.g. insufflation, face mask, cannula) are {{c1::open}} breathing systems. 
Published 12/13/2023 A circle system is a {{c2::semi-open}} breathing system when the {{c1::FGF > minute ventilation}}. 
Published 12/13/2023 A circle system is a {{c2::semi-closed}} breathing system when the {{c1::FGF < minute ventilation}}. 
Published 12/13/2023 A circle system is a {{c2::closed}} breathing system when the FGF is {{c1::very low}}. 
Published 12/13/2023 A circle system is a {{c2::closed}} breathing system when the APL valve is {{c1::closed}}. 
Published 12/13/2023 In a {{c1::closed}} breathing system, gas does not exit the scavenger. 
Published 12/13/2023 A unidirectional valve that is stuck {{c2::open}} will cause {{c1::rebreathing}}. 
Published 12/13/2023 A unidirectional valve that is stuck {{c2::closed}} will cause {{c1::airway obstruction}}. 
Published 12/13/2023 Unidirectional valves should be tested with the anesthesia gas machine in {{c1::bag and vent}} mode. 
Published 12/13/2023 Rebreathing of CO2 will cause the {{c1::β angle}} to {{c2::widen}} on the capnograph. 
Published 12/13/2023 Which class of Mapleson circuit is least efficient for spontaneous ventilation?{{c1::Mapleson B}}
Published 12/13/2023 Which class of Mapleson circuit is least efficient for controlled ventilation?{{c1::Mapleson A}}
Published 12/13/2023 The {{c1::Pethick test}} is used to evaluate unrecognized kinking of the fresh gas hose in a {{c2::Bain circuit}}. 
Published 12/13/2023 Use of a Mapleson circuit {{c1::increases}} the apparatus dead space. 
Published 12/13/2023 With a Mapleson circuit, a {{c1::decreased}} tidal volume will minimize rebreathing. 
Published 12/13/2023 With a Mapleson circuit, a {{c1::longer::shorter/longer}} expiratory time will minimize rebreathing. 
Published 12/13/2023 In a Bain system, the inhaled gas is humidified and warmed by {{c1::exhaled gas}}. 
Published 12/13/2023 What type of ventilation is a Bain circuit used for?{{c1::Spontaneous and controlled}}
Published 12/13/2023 When performing a Pethick test, if the reservoir bag {{c2::collapses}}, the Bain circuit is {{c1::safe}} to use. 
Published 12/13/2023 When performing a Pethick test, if the reservoir bag {{c2::remains inflated}}, the Bain circuit is {{c1::not safe}} to use. 