Slide quality: what causes wrinkled or folded epidermis?
Back
Occurs usually when cutting tissue across blade, picking up wafer for
slide placement, or inappropriate handling of the brush or tools
◦ Hold slide at appropriate angle to ensure proper lift and placement of
tissue onto slide – wrinkling of tissue if angle too shallow and stretching
of tissue if angle too steep
◦ High humidity in lab, cryostat cutting blade not at optimal cutting
temperature, and moisture on slide can lead to curling of tissue
slide placement, or inappropriate handling of the brush or tools
◦ Hold slide at appropriate angle to ensure proper lift and placement of
tissue onto slide – wrinkling of tissue if angle too shallow and stretching
of tissue if angle too steep
◦ High humidity in lab, cryostat cutting blade not at optimal cutting
temperature, and moisture on slide can lead to curling of tissue